Z-0884-2026 Class II Ongoing
Recalled by Illumina, Inc.
— San Diego, CA
FDA device recall Z-0884-2026 was initiated by Illumina, Inc. on Oct 28, 2025 and is designated Class II. Reason for recall: An instrument intended for sequencing DNA libraries with in vitro diagnostic assays is used with reagent kits including flow cells that have flow cell gaskets that have a quality issue, which may lead… The recall status is ongoing. Affected quantity: 11916.
Recall details
- Product Type
- Devices
- Report Date
- Dec 10, 2025
- Initiation Date
- Oct 28, 2025
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 11916
Product description
Reagent Kits/ containing flow cells, components of NextSeq 550Dx Instrument: NextSeq 550Dx High Output Reagent Kit v2.5 (75 cycles) IVD, REF: 20028870/NextSeq 550Dx HO FC Cart v2.5, 75 Cycles, REF: 20031098; NextSeq 550Dx High Output Reagent Kit v2.5 (300 cycles) IVD, REF: 20028871/NextSeq 550Dx HO FC Cart v2.5, 300 Cycles, REF: 20026365; CN NextSeq 550Dx High Output Reagent Kit v2.5 (75 cycles) IVD, REF: 20064341/CN NEXTSEQ 550Dx HO FLOW CELL V2.5, REF: 20062316; CN NextSeq 550Dx High Output Reagent Kit v2.5 (300 cycles) IVD, REF: 20064344/CN NEXTSEQ 550Dx HO FLOW CELL V2.5, REF: 20062316; CN NextSeq 550Dx Mid Output Reagent Kit v2.5 (150 cycles) IVD, 20064345/CN NEXTSEQ 550Dx MO FLOW CELL V2.5, REF: 20062530; CN NextSeq 550Dx Mid Output Reagent Kit v2.5 (300 cycles) IVD, 20064346/CN NEXTSEQ 550Dx MO FLOW CELL V2.5, REF: 20062530; /FLOWCELL, FIT FLUIDICS TESTING, REF: 15050205
Reason for recall
An instrument intended for sequencing DNA libraries with in vitro diagnostic assays is used with reagent kits including flow cells that have flow cell gaskets that have a quality issue, which may lead to flow cell leaks during runs, which could cause run failures, that may result in: No/insufficient data output, instrument fluidics clogs, toxic substance exposure to user.
Distribution pattern
Worldwide - US Nationwide distribution in the states of PA, NJ, CA, NC, UT, IL, MD, NY, VA, WV, SD, TX, AL, SC, TN, KY, MI and the countries of DK, SA, FR, DE, IT, CN, PL, GB, TH, GR, VN, BE, ES, IL, KR, AU, CH, BG, NL, IE, TW, TR, CZ, HU, NO, FI, JP, SG, LU, AT, AE, ZA, PT, LT, SE.
Code information
NextSeq 550Dx UDI-DI: 00816270020125. Kit REF/Flow Cell REF/UDI-DI(Lot): 20028870/20031098/00816270020132(A184824-3, A185016-3, A183951-3, A182816-3, A183339-3, A182533-3, A183950-3, A184531-3, A182413-3, A184058-3, A183613-3, A183211-3, A183612-3, A183826-3, A182302-3, A183449-3, A182298-3, A182532-3, A182299-3, A183830-3, A183118-3, A182301-3, A184359-3, A184727-3); 20028871/20026365/00816270020118(A182717-3, A182051-3, A179804-3, A183343-3); 20064341/20062316/006975709330014(A183402-4, A183550-4); 20064344/20062316/006975709330038(A186185-4, A183170-4, A183404-4, A183555-4, A183557-4); 20064345/20062530/006975709330045(A186074-4, A183278-4, A183405-4, A184005-4); 20064346/20062530/006975709330052(A185054-4, A183279-4, A183409-4, A183568-4, A183569-4, A183570-4, A184776-4); 15050205(20935999, 20956354)
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0884-2026
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab
- Nov 17, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.