Z-0853-2015 Class II Terminated
FDA device recall Z-0853-2015 was initiated by Zimmer, Inc. on Dec 4, 2014 and is designated Class II. Reason for recall: The affected lot was supplied by contract manufacturer and are potentially subject to disassociation of the magnet cover. Zimmer received one complaint indicating the magnet cover of a Persona… The recall status is terminated (terminated May 13, 2015). Affected quantity: 83 units.
Recall Details
- Product Type
- Devices
- Report Date
- Dec 31, 2014
- Initiation Date
- Dec 4, 2014
- Termination Date
- May 13, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 83 units
Product Description
Persona Cemented Tibial Broach Inserter/Extractor. The Persona Cemented Tibial Broach Inserter/Extractor Handle is a Use sterilizable instrument intended for use In multiple TKA procedures. The Persona Surgical Technique indicates that the Persona Cemented Tibial Broach Inserter/Extractor Handle is intended to broach the proximal tibia in preparation for the full seating of the tibial implant.
Reason for Recall
The affected lot was supplied by contract manufacturer and are potentially subject to disassociation of the magnet cover. Zimmer received one complaint indicating the magnet cover of a Persona Cemented Tibial Broach Inserter/Extractor Handle lot 62085420 was missing after broaching the tibia. Postoperative radiographs were inspected and the magnet cover was not identified. The investigation indi
Distribution Pattern
Distributed in the states of MN; NJ; NC; PA; AL; MI; OR; FL; OH; CO; IL and CA, and the countries of Australia, India, UAE, Germany, and Italy.
Code Information
Item: 42-5399-023-00 Lots: 62072307 62072565 62085420 62192754
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0853-2015
- openFDA file date
- Oct 23, 2026
- Loaded by RegDataLab