Z-0855-2022 Class II Ongoing

Recalled by Medtronic MiniMed — Northridge, CA

FDA device recall Z-0855-2022 was initiated by Medtronic MiniMed on Jan 27, 2022 and is designated Class II. Reason for recall: New or replacement insulin pumps are not pre-programmed with basal rates or other verified settings (bolus wizard settings, sensor settings, etc.), which must be set up and saved on pumps prior to… The recall status is ongoing. Affected quantity: 175,801.

Recall Details

Product Type
Devices
Report Date
Apr 6, 2022
Initiation Date
Jan 27, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
175,801

Product Description

MiniMed 630G: MMT-1715, MMT-1754, MMT-1755; MiniMed 670G: MMT-1740, MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782; MiniMed 770G: MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892; MiniMed 620G (OUS Only): MMT-1750; MiniMed 640G (OUS Only): MMT-1711, MMT-1712, MMT-1751, MMT-1752; MiniMed 720G (OUS Only): MMT-1809, MMT-1810, MMT-1859, MMT-1860; MiniMed 740G (OUS Only): MMT-1811, MMT-1812, MMT-1861, MMT-1862; MiniMed 780G (OUS Only): MMT-1884, MMT-1885, MMT-1886, MMT-1895, MMT-1896 The MiniMed" 630G System with SmartGuard is intended for continuous delivery of basal insulin (at user selected rates) and administration of insulin boluses (in user selectable amounts) for the management of Type 1 diabetes mellitus in persons sixteen years of age and older.

Reason for Recall

New or replacement insulin pumps are not pre-programmed with basal rates or other verified settings (bolus wizard settings, sensor settings, etc.), which must be set up and saved on pumps prior to use. Patients who do not realize this will receive no basal insulin, which can lead to insulin under-delivery. Insulin under-delivery can cause severe hyperglycemia, which may lead to diabetic ketoacidosis. Current device labeling does not clearly identify that these insulin pumps are capable of functioning with the default basal setting of 0.0 units/hr for all 24 hours and that no error message will appear if the default basal setting remains active. Furthermore, the labeling for the affected devices does not highlight that saving the basal settings requires multiple steps and failure to complete all steps will not save the new basal settings, nor that the steps required to set and save a basal profile and other parameters which are used as part of the bolus calculator function are different.

Distribution Pattern

US: OH, UT, IL, NY, FL, LA, NE, NV, CA, VA, TX, DE, KS, MS, MI, AL, OK, AZ, SC, IN, AR, PA, WV, MA, RI, NJ, CO, TN, KY, GA, WI, MN, NM, MD, IA, NC, WY, MO, WA, OR, CT, ND, ME, VT, ID, SD, AK, MT, NH, DC, HI, PR, VI OUS: Australia, Japan, Korea, New Zealand, Taiwan, Viet Nam

Code Information

Insulin pumps distributed in the last 6 months. All devices (US) serial numbers ranging from NG2056884H TO NG2891033H and (OUS) serial numbers ranging from NG1921788H TO NG2883091

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0855-2022
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.