Z-0863-2015 Class II Terminated

Recalled by Bausch & Lomb Inc — Saint Louis, MO

FDA device recall Z-0863-2015 was initiated by Bausch & Lomb Inc on Nov 4, 2014 and is designated Class II. Reason for recall: A software anomaly was identified where the system may automatically transition from "Standby" to "Ready to Fire" state while in surgical use. The recall status is terminated (terminated Mar 23, 2017). Affected quantity: 79.

Recall Details

Product Type
Devices
Report Date
Dec 31, 2014
Initiation Date
Nov 4, 2014
Termination Date
Mar 23, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
79

Product Description

Bausch & Lomb Stellaris PC w/ Laser Upgrade BL1433 (Stellaris PC w v3 Software) with BL2272 (Laser Upgrade kit) Stellaris PC Vision Enhancement System with integrated laser module. System is used for the surgical treatment/removal of cataractous lens, vitreous removal and certain treatments of the posterior portion of the eye.

Reason for Recall

A software anomaly was identified where the system may automatically transition from "Standby" to "Ready to Fire" state while in surgical use.

Distribution Pattern

Worldwide Distribution - USA Distribution including to the states of AL, CA, CT, FL, GA, HI, IL, IN, KS, MI, MN, MO, MS, NH, NY, OH, OK, PA, TX, WI and WV., and to the countries of Austria, Australia, Belgium, China, Czech Republic, France, Germany, Great Britain, Greece, Hong Kong, Hungary, India, Ireland, Italy, Jordan, Malaysia, Morocco, Netherlands, Norway, Portugal, Singapore, Spain, Sweden, Switzerland, Turkey and the United Arab Emirates.

Code Information

Serial numbers: SPC00146, SPC00148, SPC00181, SPC00362, SPC00384, SPC00393, SPC00403, SPC00406, SPC00418, SPC00422, SPC00424, SPC00427, SPC00466, SPC00469, SPC00500, SPC00504, SPC00524, SPC00525, SPC00566, SPC00636, SPC00738, SPC00771, SPC00785, SPC00793, SPC00802, SPC00839, SPC00851, SPC00889, SPC00926, SPC00966, SPC00969, SPC01002, SPC01035, SPC01159, SPC01167, SPC01213, SPC01218, SPC01224, SPC01263, SPC01267, SPC01300, SPC01347, SPC01405, SPC01442, SPC01446, SPC01447, SPC01449, SPC01466, SPC01474, SPC01492, SPC01542, SPC01573, SPC01608, SPC01629, SPC01653, SPC01655, SPC01666, SPC01667, SPC01678, SPC01681, SPC01687, SPC01696, SPC01699, SPC01707, SPC01732, SPC01742, SPC01746, SPC01749, SPC01750, SPC01752, SPC01754, SPC01756, SPC01760, SPC01816, SPC01817, SPC01819, SPC01820, SPC01843 and SPC01884

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0863-2015
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.