Z-0873-2022 Class II Terminated

Recalled by B Braun Medical Inc — Bethlehem, PA

Recall Details

Product Type
Devices
Report Date
April 20, 2022
Initiation Date
January 28, 2022
Termination Date
May 15, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
95

Product Description

Conductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Conductivity Sensor (END) - VERSION 2, Article Code Number 3456103A. These conductivity sensors may be used as spare parts for the following dialysis machines: 710200L (UDI 04046963686345), 710200S (UDI 04046964285608), 710200U (UDI 04046964285615), 710200K (UDI 04046963686338), 710500L (UDI Not applicable) and 710500K (UDI Not applicable).

Reason for Recall

Firm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in excessive ultrafiltration or insufficient ultrafiltration depending on whether the dialysis machine is equipped with a Dialysis Fluid filter.

Distribution Pattern

US Nationwide distribution.

Code Information

Lots 4/21 and 5/21