Z-0873-2022 Class II Terminated
FDA device recall Z-0873-2022 was initiated by B Braun Medical Inc on Jan 28, 2022 and is designated Class II. Reason for recall: Firm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in excessive ultrafiltration or… The recall status is terminated (terminated May 15, 2024). Affected quantity: 95.
Recall Details
- Product Type
- Devices
- Report Date
- Apr 20, 2022
- Initiation Date
- Jan 28, 2022
- Termination Date
- May 15, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 95
Product Description
Conductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Conductivity Sensor (END) - VERSION 2, Article Code Number 3456103A. These conductivity sensors may be used as spare parts for the following dialysis machines: 710200L (UDI 04046963686345), 710200S (UDI 04046964285608), 710200U (UDI 04046964285615), 710200K (UDI 04046963686338), 710500L (UDI Not applicable) and 710500K (UDI Not applicable).
Reason for Recall
Firm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in excessive ultrafiltration or insufficient ultrafiltration depending on whether the dialysis machine is equipped with a Dialysis Fluid filter.
Distribution Pattern
US Nationwide distribution.
Code Information
Lots 4/21 and 5/21
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0873-2022
- openFDA file date
- Oct 26, 2026
- Loaded by RegDataLab