Z-0877-2026 Class II Ongoing

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

FDA device recall Z-0877-2026 was initiated by Siemens Medical Solutions USA, Inc on Oct 24, 2025 and is designated Class II. Reason for recall: There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary… The recall status is ongoing. Affected quantity: 7 units (3 US, 4 OUS).

Recall details

Product Type
Devices
Report Date
Dec 10, 2025
Initiation Date
Oct 24, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7 units (3 US, 4 OUS)

Product description

MAMMOMAT Revelation;

Reason for recall

There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.

Distribution pattern

Worldwide - US Nationwide distribution in the states of AR, CA, NY and the countries of Australia, Canada, India.

Code information

Model Number: 11343300; UDI-DI: 04056869124575; System Serial Numbers: 1296, 10550, 10572, 10606, 10608, 11511, 20140;