Z-0885-2022 Class II Ongoing

Recalled by Volcano Corporation — Rancho Cordova, CA

Recall Details

Product Type
Devices
Report Date
April 13, 2022
Initiation Date
January 19, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3,216

Product Description

SyncVision Systems, Model: 400-0100.10, PN: 30000485688x, with software version 4.2.x

Reason for Recall

If FFR measurement(s) are made prior to an iFR/FFR co-registration/pullback in the same procedural session, this could result in incorrect iFR/FFR Co-Registration results to be displayed, that may cause users to mistakenly use incorrect measurements, leading to inappropriate patient treatment. Image acquisition and processing system operators manual to be updated to include workflow alternatives.

Distribution Pattern

US: WI, CA, CO, MO, ID, DC, LA, VA, KS, OK, GA, KY, IL, AZ, NY, IN, FL, TX, NC, OH, WA, OR, NM, MT, IA, MI, TN, MN, PA, AL, MA, WV, NV, NH, CT, NE, NJ, HI, AR, UT, MD, SC, MS, AK, DE OUS: JP, SA, GB, DE, PT, SG, PL, ZA, NL, MT, ES, IN, GR, EG, FI, HK, FR, NO, DK, BE, AT, IT, CO, KW, MX, PS, QA, CY, CA, TW, LV, SE, HR, SI, MY, KR, IL, TH, AU, BR, CL, SK, CH, LB, OM, ID, PK, IE, NZ, PR, RO, MM, AR, RS, RU, PA, KE, UZ

Code Information

All serial numbers, limited to cases when: The SyncVision runs the current supported software version 4.2.x, AND The secondary modality (IntraSight IVUS) runs software version 5.x. Device Identifier (DI): 00845225012434