Z-1891-2018 Class II Terminated

Recalled by Volcano Corporation — Rancho Cordova, CA

Recall Details

Product Type
Devices
Report Date
May 23, 2018
Initiation Date
March 22, 2018
Termination Date
April 28, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
8 devices

Product Description

Philips Volcano CORE M2 Vascular System, Part # 400-0100.17, Rx Only. Used for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature.

Reason for Recall

The incorrect resistor has been used in affected CORE M2 Imaging Systems, which causes noise and results in an overly bright IVUS Image.

Distribution Pattern

Worldwide Distribution -- U.S., to the states of California, Connecticut, Tennessee, North Carolina, Maryland, and Oregon; and, the country of Australia.

Code Information

Serial #: 126620101597, 126620101599, 126620101600, 126620100718, 126620102313, 126620102315, 126620102307, and 126620093077.