Z-0892-2017 Class II Terminated
FDA device recall Z-0892-2017 was initiated by Medtronic Inc on Nov 10, 2016 and is designated Class II. Reason for recall: Medtronic identified an issue with specific models and lots of the PowerCross .018 Over the Wire (OTW) Percutaneous Transluminal Angioplasty (PTA) Dilatation Catheter. During an unrelated… The recall status is terminated (terminated Aug 2, 2017). Affected quantity: 315.
Recall Details
- Product Type
- Devices
- Report Date
- Dec 28, 2016
- Initiation Date
- Nov 10, 2016
- Termination Date
- Aug 2, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 315
Product Description
ev3, PowerCross .018 OTW PTA Dilatation Catheter, Sterile EO. The PowerCross .018 OTW PTA Dilatation Catheter is intended to dilate stenoses in the iliac, femoral, ilio-femoral, popliteal, infrapopliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent post-dilatation in the peripheral vasculature.
Reason for Recall
Medtronic identified an issue with specific models and lots of the PowerCross .018 Over the Wire (OTW) Percutaneous Transluminal Angioplasty (PTA) Dilatation Catheter. During an unrelated manufacturing inspection of scrapped products, an out of specification condition was observed. The catheter outer shafts were pinched and deformed at the proximal balloon bond. Bench testing of affected devices showed that this resulted in either the inability to inflate the device, a slow deflation, or the inability to deflate the device post expansion.
Distribution Pattern
NY, NC, WA
Code Information
UDI/Product Lot/Serial Product Number UPN/GTIN Number 00821684046647 A358626 AB18W030100150 00821684046883 A357386 AB18W060100150 00821684046722 A357070 AB18W040100150 00821684046777 A358043 AB18W050040150 00821684046722 A356360 AB18W040100150 00821684046777 A357348 AB18W050040150 00821684046562 A359219 AB18W025040150
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0892-2017
- openFDA file date
- Oct 23, 2026
- Loaded by RegDataLab