Z-0900-2015 Class II Terminated

Recalled by Zimmer, Inc. — Warsaw, IN

FDA device recall Z-0900-2015 was initiated by Zimmer, Inc. on Dec 8, 2014 and is designated Class II. Reason for recall: Samples of product from Supplier lot 14D0809 and 14D0808 showed little adhesive transfer between the Tyvek lid and cavity when opened for internal bone cement cure and pairing tests. Visually these… The recall status is terminated (terminated Jul 17, 2015). Affected quantity: 40 units.

Recall Details

Product Type
Devices
Report Date
Jan 7, 2015
Initiation Date
Dec 8, 2014
Termination Date
Jul 17, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
40 units

Product Description

Osteobond Bone Cement Bone cement monomer is kitted with bone cement powder to create a cement mantle for orthopedic implants.

Reason for Recall

Samples of product from Supplier lot 14D0809 and 14D0808 showed little adhesive transfer between the Tyvek lid and cavity when opened for internal bone cement cure and pairing tests. Visually these samples do not meet the requirements of ZWI 43.109 and corresponding visual aid 55-0000-310-01.

Distribution Pattern

Distributed to MI, KS, and CA.

Code Information

PMA #N17755, decision date 5/17/1976, reclassified 10/14/1999. Class II device; D033511, Item No. 00-1101-012-00; lot 62849791, 14D0809, Manufactured between 9/11/2014-10/6/2014; EXP date: 4/30/2017

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0900-2015
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.