Z-0902-2015 Class II Terminated
FDA device recall Z-0902-2015 was initiated by Horiba Instruments, Inc Dba Horiba Medical on Nov 13, 2014 and is designated Class II. Reason for recall: HORIBA Medical is recalling the ABX PENTRA 400 Reagent Container because they do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent… The recall status is terminated (terminated Mar 19, 2015). Affected quantity: 443 units.
Recall details
- Product Type
- Devices
- Report Date
- Jan 14, 2015
- Initiation Date
- Nov 13, 2014
- Termination Date
- Mar 19, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 443 units
Product description
ABX PENTRA Reagent Container, Model No. B1037307 (15 mL), B1034626 (10 mL), and B1034634 (4 mL)
Reason for recall
HORIBA Medical is recalling the ABX PENTRA 400 Reagent Container because they do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm. The defect would result in a false high or false low results for HbA1c and false high results for magnes
Distribution pattern
Nationwide in US
Code information
Model No. B1037307 (15 mL), B1034626 (10 mL), and B1034634 (4 mL) manufactured since June 2014.