Z-0902-2015 Class II Terminated

Recalled by Horiba Instruments, Inc dba Horiba Medical — Irvine, CA

Recall Details

Product Type
Devices
Report Date
January 14, 2015
Initiation Date
November 13, 2014
Termination Date
March 19, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
443 units

Product Description

ABX PENTRA Reagent Container, Model No. B1037307 (15 mL), B1034626 (10 mL), and B1034634 (4 mL)

Reason for Recall

HORIBA Medical is recalling the ABX PENTRA 400 Reagent Container because they do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm. The defect would result in a false high or false low results for HbA1c and false high results for magnes

Distribution Pattern

Nationwide in US

Code Information

Model No. B1037307 (15 mL), B1034626 (10 mL), and B1034634 (4 mL) manufactured since June 2014.