Z-0902-2015 Class II Terminated

Recalled by Horiba Instruments, Inc Dba Horiba Medical — Irvine, CA

FDA device recall Z-0902-2015 was initiated by Horiba Instruments, Inc Dba Horiba Medical on Nov 13, 2014 and is designated Class II. Reason for recall: HORIBA Medical is recalling the ABX PENTRA 400 Reagent Container because they do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent… The recall status is terminated (terminated Mar 19, 2015). Affected quantity: 443 units.

Recall details

Product Type
Devices
Report Date
Jan 14, 2015
Initiation Date
Nov 13, 2014
Termination Date
Mar 19, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
443 units

Product description

ABX PENTRA Reagent Container, Model No. B1037307 (15 mL), B1034626 (10 mL), and B1034634 (4 mL)

Reason for recall

HORIBA Medical is recalling the ABX PENTRA 400 Reagent Container because they do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm. The defect would result in a false high or false low results for HbA1c and false high results for magnes

Distribution pattern

Nationwide in US

Code information

Model No. B1037307 (15 mL), B1034626 (10 mL), and B1034634 (4 mL) manufactured since June 2014.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0902-2015
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0902-2015. Report a data issue: hello@regdatalab.com.