Z-0924-2014 Class III Terminated

Recalled by Horiba Instruments, Inc Dba Horiba Medical — Irvine, CA

FDA device recall Z-0924-2014 was initiated by Horiba Instruments, Inc Dba Horiba Medical on Dec 18, 2013 and is designated Class III. Reason for recall: HORIBA Medical is informing all ABX PENTRA Urine Control L/H customers running the Creatinine assay on ABX PENTRA Chemistry Analyzers that the Target Range for Creatinine has been updated. The ABX… The recall status is terminated (terminated Oct 9, 2014). Affected quantity: 531.

Recall details

Product Type
Devices
Report Date
Feb 12, 2014
Initiation Date
Dec 18, 2013
Termination Date
Oct 9, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
531

Product description

ABX PENTRA Urine Control L/H. The ABX PENTRA Urine Control L/H is a two-level (Low and High) quality control consisting of liquid solutions prepared from human urine with chemical additives and materials of biological origin added as required to obtain given component levels. The assigned values of the control components are given in the labeling, ensuring control of the appropriate HORIBA ABX SAS methods on the ABX PENTRA chemistry analyzers. Each control level is provided in one vial of 10 ml. The ABX PENTRA Urine Control L/H is for use in quality control by monitoring accuracy and precision.

Reason for recall

HORIBA Medical is informing all ABX PENTRA Urine Control L/H customers running the Creatinine assay on ABX PENTRA Chemistry Analyzers that the Target Range for Creatinine has been updated. The ABX PENTRA Urine Control L/H lot specific annexes have been updated to include revised acceptable Target Ranges for Creatinine.No other analyte values are affected.

Distribution pattern

Worldwide Distribution - USA (nationwide) and Canada.

Code information

Model/Part #(s): A11A01674. Lot/Serial #(s): 1207001, 1303601, and 1300701. Shelf Life: Lot#s 1207001 & 1300701 Expire on: 06/30/2014. Lot# 1303601 Expires on: 09/30/2014.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0924-2014
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0924-2014. Report a data issue: hello@regdatalab.com.