Z-0906-2022 Class II Ongoing

Recalled by Integra LifeSciences Corp. — Princeton, NJ

FDA device recall Z-0906-2022 was initiated by Integra LifeSciences Corp. on Feb 28, 2022 and is designated Class II. Reason for recall: Excess force used to tighten the Universal Flexible Arm can cause the internal cable to break at the tip, resulting in an unusable condition. The recall status is ongoing. Affected quantity: 54.

Recall Details

Product Type
Devices
Report Date
Apr 20, 2022
Initiation Date
Feb 28, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
54

Product Description

Integra Universal Flexible Arm part number REF 1362275

Reason for Recall

Excess force used to tighten the Universal Flexible Arm can cause the internal cable to break at the tip, resulting in an unusable condition.

Distribution Pattern

US Nationwide Distribution: AL, AZ, FL, GA, IL, MA, MD, NY, OH, OR, PA, TX

Code Information

UDI#10381780074670, Lots: AC2109, AC2103, AC2106, AC2110, AC2101, AC2008, AC2009

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0906-2022
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.