Z-0925-2013 Class II Terminated

Recalled by Biomet, Inc. — Warsaw, IN

FDA device recall Z-0925-2013 was initiated by Biomet, Inc. on Feb 5, 2013 and is designated Class II. Reason for recall: The Straight Ringloc Button Latch Inserter Handle may have been manufactured at the higher end of the tolerance causing interference with the Ringloc cup. The inserter handle may lock onto the cup… The recall status is terminated (terminated Sep 30, 2013). Affected quantity: 58.

Recall Details

Product Type
Devices
Report Date
Mar 20, 2013
Initiation Date
Feb 5, 2013
Termination Date
Sep 30, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
58

Product Description

REF S313152 Straight Acetabular Inserter Handle Ringloc non-sterile Orthopedic manual surgical instrument

Reason for Recall

The Straight Ringloc Button Latch Inserter Handle may have been manufactured at the higher end of the tolerance causing interference with the Ringloc cup. The inserter handle may lock onto the cup during impaction. If alternative inserters are not available this could cause a surgical delay greater than 30 minutes.

Distribution Pattern

Worldwide Distribution - No USA distribution -Internationally to Canada, Japan, and Chile

Code Information

Catalog Number 5313152 Lot 153880,153930, 415500, 648010