Z-0928-2022 Class II Ongoing

Recalled by Siemens Healthcare Diagnostics, Inc. — Tarrytown, NY

Recall Details

Product Type
Devices
Report Date
April 20, 2022
Initiation Date
February 9, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2,395

Product Description

Atellica CH Calcium_2 (CA_2), Siemens Material Number 11097644

Reason for Recall

Siemens Healthcare Diagnostics Inc. has confirmed the potential for reagent carryover impacting quality control (QC), patient samples, and calibrator results.

Distribution Pattern

Global Distribution. US Nationwide.

Code Information

UDI: 00630414220697; Lot Numbers: 211392 211438 211484 211522 211526