Z-0947-2019 Class II Terminated

Recalled by Arrow International Inc — Reading, PA

Recall Details

Product Type
Devices
Report Date
March 6, 2019
Initiation Date
October 2, 2018
Termination Date
November 12, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5

Product Description

AF 550 x 5ML/H FH DL 30hole SAT C,4"IN, Product code MC0050XLYK2 CP, Arrow AutoFuser Disposable Pain Control Pump

Reason for Recall

Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.

Distribution Pattern

Nationwide distribution.

Code Information

74B1700141 74K1602023 74J1601147 74L1602735 74J1602863