Z-0949-2016 Class II Terminated

Recalled by Xeridiem Mediem Medical Devices Inc — Tucson, AZ

FDA device recall Z-0949-2016 was initiated by Xeridiem Mediem Medical Devices Inc on Dec 23, 2015 and is designated Class II. Reason for recall: A feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus… The recall status is terminated (terminated Oct 25, 2016). Affected quantity: 154.

Recall Details

Product Type
Devices
Report Date
Mar 2, 2016
Initiation Date
Dec 23, 2015
Termination Date
Oct 25, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
154

Product Description

EndoVive 3s Low Profile Balloon Kits Part Number: M00548550 (XMD P/N 70-0050-616) The EndoVive 3s Low Profile Balloon is intended to provide gastric access through a gastrointestinal stoma tract for enteral feeding, medication administration and gastric decompression. The EndoVive 3s Low Profile Balloon is indicated for use in adult and pediatric patients who require long-term feeding, are unable to tolerate oral feeding, are at low risk for aspiration or require gastric decompression and/or medication delivery directly into the stomach.

Reason for Recall

A feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric contents to leak out of the device in some instances.

Distribution Pattern

Nationwide distribution to MA only.

Code Information

Part Number: M00548550 (XMD P/N 70-0050-616) Lot Number: 928917 929625

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0949-2016
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.