Z-0952-2026 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- December 24, 2025
- Initiation Date
- November 7, 2025
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 630 units
Product Description
Brand Name: Rocket Product Name: Rocket 18Fg Seldinger Catheter Procedure Tray Model/Catalog Number: R54544-18-PK Software Version: N/A Product Description: Chest tube manufactured from PVC, with fenestrations and a barium stripe to the indicated French gauge size. Supplied with a procedure pack to allow percutaneous insertion, supplied with connectors to allow attachment to chest drain bottle devices. Component: N/A
Reason for Recall
IFU and device kit labeling contains misleading statements leading to improper securing of the device resulting in the potential to fall out during use.
Distribution Pattern
Domestic: CO, CT, IL, MA, ME, NH, NJ, NM, OH, RI, SC, TX; International: UAE, Austria, Australia, Canada, Germany, Denmark, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, India, Italy, Jersey, Kenya, Sri Lanka, Malta, Netherlands, Norway, New Zealand, Portugal, Saudi Arabia, Singapore, Slovenia, South Africa
Code Information
Model/Catalog Number: R54544-18-PK; UDI-DI: 050552709TF06DXY; Lot Code: 496899, 500956, 497076, 501169, 497584, 501188, 499227, 502430, 499417, 502951, 499461, 503306, 499897, 503430, 499898, 503529, 499899, 504180, 499908, 504543, 500407, 504823, 500477;