Z-0953-2026 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- December 24, 2025
- Initiation Date
- November 7, 2025
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 110 units
Product Description
Brand Name: Rocket Product Name: Rocket 18Fg Seldinger Catheter with Safety Guard Model/Catalog Number: R54544-18-SG Software Version: N/A Product Description: Chest tube manufactured from PVC, with fenestrations and a barium stripe to the indicated French gauge size. Supplied with a procedure pack to allow percutaneous insertion, supplied with connectors to allow attachment to chest drain bottle devices.
Reason for Recall
IFU and device kit labeling contains misleading statements leading to improper securing of the device resulting in the potential to fall out during use.
Distribution Pattern
Domestic: CO, CT, IL, MA, ME, NH, NJ, NM, OH, RI, SC, TX; International: UAE, Austria, Australia, Canada, Germany, Denmark, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, India, Italy, Jersey, Kenya, Sri Lanka, Malta, Netherlands, Norway, New Zealand, Portugal, Saudi Arabia, Singapore, Slovenia, South Africa
Code Information
Model/Catalog Number: R54544-18-SG; UDI-DI: 050552709TF06DXY; Lot Code: 496809, 501616, 497346, 501710, 497665, 502431, 498972, 503735, 499909, 503937, 500782, 504573, 501256, 504824;