Z-0953-2026 Class II Ongoing

Recalled by Rocket Medical Plc — Tyne & Wear

Recall Details

Product Type
Devices
Report Date
December 24, 2025
Initiation Date
November 7, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
110 units

Product Description

Brand Name: Rocket Product Name: Rocket 18Fg Seldinger Catheter with Safety Guard Model/Catalog Number: R54544-18-SG Software Version: N/A Product Description: Chest tube manufactured from PVC, with fenestrations and a barium stripe to the indicated French gauge size. Supplied with a procedure pack to allow percutaneous insertion, supplied with connectors to allow attachment to chest drain bottle devices.

Reason for Recall

IFU and device kit labeling contains misleading statements leading to improper securing of the device resulting in the potential to fall out during use.

Distribution Pattern

Domestic: CO, CT, IL, MA, ME, NH, NJ, NM, OH, RI, SC, TX; International: UAE, Austria, Australia, Canada, Germany, Denmark, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, India, Italy, Jersey, Kenya, Sri Lanka, Malta, Netherlands, Norway, New Zealand, Portugal, Saudi Arabia, Singapore, Slovenia, South Africa

Code Information

Model/Catalog Number: R54544-18-SG; UDI-DI: 050552709TF06DXY; Lot Code: 496809, 501616, 497346, 501710, 497665, 502431, 498972, 503735, 499909, 503937, 500782, 504573, 501256, 504824;