Z-0956-2019 Class II Terminated
FDA device recall Z-0956-2019 was initiated by Arrow International Inc on Oct 2, 2018 and is designated Class II. Reason for recall: Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is… The recall status is terminated (terminated Nov 12, 2021). Affected quantity: 1.
Recall Details
- Product Type
- Devices
- Report Date
- Mar 6, 2019
- Initiation Date
- Oct 2, 2018
- Termination Date
- Nov 12, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1
Product Description
AF w/AS 550 x 1-14ML/HR 5ML-60LO CPNB, Product code MVBX60XL CPNB, Arrow AutoFuser Disposable Pain Control Pump
Reason for Recall
Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.
Distribution Pattern
Nationwide distribution.
Code Information
74E1702443
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0956-2019
- openFDA file date
- Oct 23, 2026
- Loaded by RegDataLab