Z-0965-2026 Class II Ongoing

Recalled by ICU Medical, Inc. — Lake Forest, IL

Recall Details

Product Type
Devices
Report Date
December 31, 2025
Initiation Date
November 24, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7 customers

Product Description

LifeShield Drug Library Management (DLM); LifeShield Infusion Safety Software Suite v2.2 List Number 17003-02

Reason for Recall

For certain customers, the Change Summary review option ceases to operate as expected for any migrated library after the migration has been completed. Users must use alternate methods to review changes. Failure to do so may lead in over delivery, under delivery, or delay in setup on the infusion pump.

Distribution Pattern

US Nationwide distribution in the states of CA, IA, LA, NY, PA, TN, VA.

Code Information

version 2.2.1 UDI-DI 00887709127227 Issue affects customers who migrated from LifeShield DLM v2.1x to LifeShield DLM v2.2.1 and have any version of a Plum Duo v1.1 drug library that was approved or locked for review before the migration.