Z-1119-2026 Class II Ongoing

Recalled by ICU Medical, Inc. — Lake Forest, IL

FDA device recall Z-1119-2026 was initiated by ICU Medical, Inc. on Dec 15, 2025 and is designated Class II. Reason for recall: Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary… The recall status is ongoing.

Recall Details

Product Type
Devices
Report Date
Jan 28, 2026
Initiation Date
Dec 15, 2025
Voluntary/Mandated
Voluntary: Firm initiated

Product Description

ICU Medical Plum Duo Precision IV Pump, List Numbers: 400020403 (US), 400021003 (Philippines)

Reason for Recall

Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.

Distribution Pattern

Worldwide distribution - US Nationwide and the country of Philippines.

Code Information

400020403 (US), Software Versions: 1.2.2; 400021003 (Philippines), Software Versions: 1.2.0

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1119-2026
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.