Z-0969-2014 Class II Terminated

Recalled by Boston Scientific Corporation — Maple Grove, MN

FDA device recall Z-0969-2014 was initiated by Boston Scientific Corporation on Feb 4, 2014 and is designated Class II. Reason for recall: Boston Scientific is initiating this Recall Removal of one lot of the Coyote" OVER-THE-WIRE PTA Balloon Dilatation Catheter and one lot of the Coyote" MONORAIL" PTA Balloon Dilatation Catheter. Boston… The recall status is terminated (terminated Aug 4, 2014). Affected quantity: 35 for both products.

Recall Details

Product Type
Devices
Report Date
Feb 19, 2014
Initiation Date
Feb 4, 2014
Termination Date
Aug 4, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
35 for both products.

Product Description

Boston Scientific, 2.5mm Coyote" OVER-THE-WIRE PTA Balloon Dilation Catheter, UPN# H74939186251210, Catalog # 39186-25121.

Reason for Recall

Boston Scientific is initiating this Recall Removal of one lot of the Coyote" OVER-THE-WIRE PTA Balloon Dilatation Catheter and one lot of the Coyote" MONORAIL" PTA Balloon Dilatation Catheter. Boston Scientific has become aware that the balloons of the devices from these lots may not deflate as expected during a procedure. To date, Boston Scientific has received three complaints for this issue. T

Distribution Pattern

US: AL, CA, FL, IL, IN, MI, MO, NY, OH, PA, TX. OUS: ITALY, GERMANY, FRANCE, GREAT BRITAIN, NETHERLANDS, WHITE RUSSIA, INDIA, JAPAN.

Code Information

16397693

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0969-2014
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026