Z-0972-2022 Class II Ongoing

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

Recall Details

Product Type
Devices
Report Date
April 27, 2022
Initiation Date
February 23, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7 systems

Product Description

Artis zee biplane MN, Model Number 10094143

Reason for Recall

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Distribution Pattern

US Nationwide.

Code Information

UDI: 4056869010106; Serial Numbers: 149100 149401 149421 149106 149202 149420 149402