Z-0973-2024 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- February 14, 2024
- Initiation Date
- January 6, 2022
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1
Product Description
Ingenia 1.5T (Product Number 781396) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system
Reason for Recall
Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.
Distribution Pattern
Worldwide - US Nationwide distribution in the states of FL and NY. The countries of Denmark, Norway, and Sweden.
Code Information
UDI-DI 00884838009820 Ingenia Serial number 41045 Accessory Serial Number 100018