Z-0974-2022 Class II Ongoing

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

Recall Details

Product Type
Devices
Report Date
April 27, 2022
Initiation Date
February 23, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
80 systems

Product Description

Artis Q floor, Model Number 10848280

Reason for Recall

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Distribution Pattern

US Nationwide.

Code Information

UDI: 4056869009971; Serial Numbers: 103221 103217 103218 103345 103409 103340 103439 104007 103302 103232 104016 104005 103391 103404 104027 132002 103334 103354 103359 103030 103335 104012 103220 103013 103384 103035 104029 104020 103039 103050 103365 103279 104025 103285 103286 103207 103222 103230 103295 103310 103430 103432 104047 103001 103239 103353 104013 103444 132502 132500 103048 103315 103263 103356 132000 103028 103241 103293 103388 103389 132003 103245 103348 103253 Additional Serial Numbers as of 10/4/23: 160501 160519 160406 160900 160850 160959 160461 103053 103040 160903 126206 160905 160930 160967 160925 160469