Z-0975-2024 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- February 14, 2024
- Initiation Date
- January 6, 2022
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 10
Product Description
Ingenia Ambition X (Product Number 781356) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system
Reason for Recall
Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.
Distribution Pattern
Worldwide - US Nationwide distribution in the states of FL and NY. The countries of Denmark, Norway, and Sweden.
Code Information
UDI-DI 00884838090040 Ingenia Serial number / Accessory Serial Number 47004 100020 47019 100024 47190 100022 47200 100025 47210 100019 47227 100063 47231 100056 47271 100032 47281 100028 47281 100031