Z-0978-2022 Class II Ongoing

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

FDA device recall Z-0978-2022 was initiated by Siemens Medical Solutions USA, Inc on Feb 23, 2022 and is designated Class II. Reason for recall: If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message… The recall status is ongoing. Affected quantity: 33 systems.

Recall Details

Product Type
Devices
Report Date
Apr 27, 2022
Initiation Date
Feb 23, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
33 systems

Product Description

Artis Q.zen floor, Model Number 10848353

Reason for Recall

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Distribution Pattern

US Nationwide.

Code Information

UDI: 4056869010014; Serial Numbers: 105120 105110 105033 105032 105500 105105 105028 105503 105018 105114 105115 105026 105025 105135 105016 105119 105137 105501 105132 105128 105142 105017 105121 105122 105123 105100 105112 105124 105029 105030 105031 105125 Additional Serial Numbers as of 10/4/23: 143500

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0978-2022
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.