Z-0978-2023 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- January 25, 2023
- Initiation Date
- December 14, 2022
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 812 units
Product Description
Cardinal Health Jackson-Pratt Channel Drain, 10 mm, REF JP-2215
Reason for Recall
Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal Health Jackson-Pratt Channel Drains including lack of radiopacity in wound drains, dull trocars impacting product performance, lack of flexibility of wound drains, and broken wound tips.
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Canada, Saudi Arabia, Singapore, Hong Kong.
Code Information
UDI/DI 50885380182523(cs), 20885380182522(ea), Lot Numbers: N210278, N210368, N210423, N210460, N210468 , N210469, N210543, N220050, N220099, N220139, N220163