Z-0979-2022 Class II Ongoing

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

Recall Details

Product Type
Devices
Report Date
April 27, 2022
Initiation Date
February 23, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
81 systems

Product Description

Artis Q.zen ceiling, Model Number 10848354

Reason for Recall

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Distribution Pattern

US Nationwide.

Code Information

UDI: 4056869010021; Serial Numbers: 111160 111162 111163 111137 111502 111139 111174 111185 111503 111005 111002 111145 111176 111510 111155 111179 111140 111507 111122 111168 111170 111173 111512 111141 111508 111521 111133 111003 111111 111103 111101 111011 111182 111184 111143 111150 111164 111183 111504 111149 111109 111151 111129 111121 111130 111517 111513 111017 111010 111115 111116 111117 111119 111104 111148 111127 111175 111506 111112 111131 111132 111134 111158 111167 111157 111161 111142 111156 111500 111501 111169 111505 111166 111107 111514 111000 111106 111123 111128 111135 Additional Serial Numbers as of 10/4/23: 111125