Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

FDA device recall Z-0986-2022 was initiated by Siemens Medical Solutions USA, Inc on Feb 25, 2022 and is designated Class II. Reason for recall: Five potential software issues affecting Artis pheno and Artis icono systems in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). The potential software errors are related… The recall status is ongoing. Affected quantity: 35 units.

Recall Details

Product Type
Devices
Report Date
Apr 27, 2022
Initiation Date
Feb 25, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
35 units

Product Description

Artis icono floor in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). Intended use single and biplane diagnostic imaging and interventional procedures. Model: 11327700

Reason for Recall

Five potential software issues affecting Artis pheno and Artis icono systems in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). The potential software errors are related to the block movements function and manual rotation of the flat detector. Depending on the status of the intervention, the limited functionality may not be sufficient to continue with treatment as planned. This may result in a situation where clinical treatment may be delayed until an additional restart is performed

Distribution Pattern

US Nationwide distribution.

Code Information

S/N: Serial 170319 170026 170343 170367 170368 170305 170360 170307 170331 170349 170309 170310 170361 170351 170383 170023 170304 170328 170334 170336 170025 170346 170051 170358 170357 170333 170365 170332 170312 170027 170303 170378 170322 170053 170364 UDI: 04056869149325

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0986-2022
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.