Z-0987-2022 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- April 27, 2022
- Initiation Date
- February 25, 2022
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 20 units
Product Description
Artis pheno- in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). Intended use-single and biplane diagnostic imaging and interventional procedures. Model: 10849000
Reason for Recall
Five potential software issues affecting Artis pheno and Artis icono systems in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). The potential software errors are related to the block movements function and manual rotation of the flat detector. Depending on the status of the intervention, the limited functionality may not be sufficient to continue with treatment as planned. This may result in a situation where clinical treatment may be delayed until an additional restart is performed
Distribution Pattern
US Nationwide distribution.
Code Information
Serial Number: 164719 164712 164740 164722 164763 164735 164723 164743 164741 164728 164711 164718 164739 164703 164721 164726 164724 164725 164760 UDI: 04056869046877