Z-0993-2022 Class II Ongoing

Recalled by Baxter Healthcare Corporation — Deerfield, IL

FDA device recall Z-0993-2022 was initiated by Baxter Healthcare Corporation on Mar 15, 2022 and is designated Class II. Reason for recall: This correction is due to software anomalies occurring during use. If the operator initiates therapy by entering a value other than the default value for the patient Gain/Loss Limit or the Return… The recall status is ongoing. Affected quantity: 3513 units.

Recall Details

Product Type
Devices
Report Date
May 4, 2022
Initiation Date
Mar 15, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3513 units

Product Description

PrisMax System, Product Codes: a) 955558, b) 955627, c) 955701

Reason for Recall

This correction is due to software anomalies occurring during use. If the operator initiates therapy by entering a value other than the default value for the patient Gain/Loss Limit or the Return Pressure Drop Limit (RDL) and uses the Same Patient function, the PrisMax System will return the patient Gain/Loss Limit or the RDL to the default value, rather than the non-default value entered at the initiation of treatment. The PrisMax System interface will display the original values once treatment is restarted but does not have an alert to inform the user of the change. The RDL value will be returned to the original value immediately upon selecting the Same Patient function. The Gain/Loss Limit value will be returned to the original value when the patient weight and/or hematocrit is changed during treatment.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Mexico, Brazil, Jamaica, Trinidad and Tobago, Canada, Australia, Singapore, Hong Kong, Vietnam, New Zealand, Korea, Austria, Belgium, Denmark, Estonia, Finland, France, Germany, Greece, Ireland, Israel, Italy, Luxemburg, Netherlands, Norway, Poland, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, Turkey, and United Kingdom.

Code Information

a) 955558, UDI 07332414124359, b) 955627, UDI 07332414124700, c) 955701, UDI 07332414125844, All Serial Numbers

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0993-2022
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.