Z-0994-2022 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- May 4, 2022
- Initiation Date
- March 7, 2022
- Termination Date
- December 19, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1 (US)
Product Description
syngo.via RT Image Suite upgraded to VB60 running on syngo.via Model Number 10496180. Running on servers with syngo.via VB60 software after upgrade and CT scanners with syngo.CT VA40A after upgrade.
Reason for Recall
Upgraded software version makes an automatic change in laser configuration settings for "markerless workflow" which may lead to the wrong laser offset coordinates being displayed in the Patient Marking step.
Distribution Pattern
Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, IL, IN, KY, MD, MO, ND, NH, NY, OH, OR, PA, SD, TX, VA and the countries of Argentina, Austria, Bahrain, Belgium, Brazil, Czechia, Denmark, France, Germany, Hungary, India, Italy, Japan, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Switzerland, United Kingdom.
Code Information
Serial Number 133527