Z-0996-2013 Class II Terminated

Recalled by Smiths Medical ASD, Inc. — Saint Paul, MN

FDA device recall Z-0996-2013 was initiated by Smiths Medical ASD, Inc. on May 14, 2010 and is designated Class II. Reason for recall: An issue with the plunger rod seal was identified where the seal had the potential to migrate inside the pump case top during slow rate/ high back pressure infusions. The recall status is terminated (terminated Apr 11, 2013). Affected quantity: 6.

Recall Details

Product Type
Devices
Report Date
Apr 3, 2013
Initiation Date
May 14, 2010
Termination Date
Apr 11, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6

Product Description

Medfusion Model 3500 Syringe Pump, Model #3500-415 and 3500-500. Intended for use in critical care areas for the administration of fluids requiring precisely controlled infusion rates.

Reason for Recall

An issue with the plunger rod seal was identified where the seal had the potential to migrate inside the pump case top during slow rate/ high back pressure infusions.

Distribution Pattern

Distributed in the state of NH.

Code Information

Lot/Serial Numbers: M76369, M76370, M76373, M76374, M76375, and M76377.