Z-0996-2022 Class II Terminated

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

Recall Details

Product Type
Devices
Report Date
May 4, 2022
Initiation Date
March 7, 2022
Termination Date
December 19, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
15 (US)

Product Description

syngo.via RT Image Suite upgraded to VB60 running on SOMATOM go.Sim Model Number 11061660. Running on servers with syngo.via VB60 software after upgrade and CT scanners with syngo.CT VA40A after upgrade.

Reason for Recall

Upgraded software version makes an automatic change in laser configuration settings for "markerless workflow" which may lead to the wrong laser offset coordinates being displayed in the Patient Marking step.

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, IL, IN, KY, MD, MO, ND, NH, NY, OH, OR, PA, SD, TX, VA and the countries of Argentina, Austria, Bahrain, Belgium, Brazil, Czechia, Denmark, France, Germany, Hungary, India, Italy, Japan, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Switzerland, United Kingdom.

Code Information

Partial UDI: 04056869249230, Serial Numbers 128123, 128106, 128109, 128144, 128124, 128112, 128132, 128135, 128114, 128125, 128113, 128142, 128107, 128136, 128133