Z-1007-2021 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- February 17, 2021
- Initiation Date
- January 4, 2021
- Termination Date
- March 11, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 7,400,000 (globally); 260,395 (US)
Product Description
INJECTOR FORCEMAX LOWER 25G 6MM, model no. NM-400U-0625 - Product Usage: Used with endoscopes to perform endoscopic injection for the treatment of esophageal and gastric varices and for submucosal injection in the digestive tract.
Reason for Recall
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Distribution Pattern
US Nationwide distribution.
Code Information
01K, 03K, 04K, 05K, 06K, 07K, 08K, 09K, 98K, 99K, 9XK, 9YK, 9ZK, 05V, 06V, 07V, 08V, 09V, 0XV