Z-1011-2024 Class II Ongoing

Recalled by Becton Dickinson & Co. — Sparks, MD

FDA device recall Z-1011-2024 was initiated by Becton Dickinson & Co. on Jan 8, 2024 and is designated Class II. Reason for recall: There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional… The recall status is ongoing. Affected quantity: 2,363,168 total units.

Recall Details

Product Type
Devices
Report Date
Feb 14, 2024
Initiation Date
Jan 8, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2,363,168 total units

Product Description

Sensi Disc Amoxicillin with Clavulanic Acid 20/10 ¿g, Catalog No. 231628 and 231629.¿ These discs are used for semi-quantitative in vitro susceptibility testing

Reason for Recall

There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.

Distribution Pattern

Domestic distribution nationwide. International distribution worldwide.

Code Information

Catalog No. 231628 UDI-DI 00382902316282 Lots 2090490 2146520 2210201 2276143 3062095; Catalog No. 231629¿¿ UDI-DI 30382902316290 Lots 3163306 3163307 3163310 3254307 3254309 3271840 3271853 2146520 2146526 2147225 2210201 2224295 2276143 2276144 3003894 3062095 3090649

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1011-2024
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 24, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.