Z-1014-2019 Class II Terminated

Recalled by Siemens Healthcare Diagnostics, Inc. — Newark, DE

FDA device recall Z-1014-2019 was initiated by Siemens Healthcare Diagnostics, Inc. on Dec 12, 2018 and is designated Class II. Reason for recall: When the Millipore Water Purification Module Progard Pretreatment Packs are used with Dimension Vista Triglyceride assay, patient and QC triglyceride results may be erroneously elevated as a result of… The recall status is terminated (terminated Apr 27, 2020). Affected quantity: 5373.

Recall Details

Product Type
Devices
Report Date
Mar 27, 2019
Initiation Date
Dec 12, 2018
Termination Date
Apr 27, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5373

Product Description

Dimension Vista 500, Analyzer, chemistry Model no. 10284473

Reason for Recall

When the Millipore Water Purification Module Progard Pretreatment Packs are used with Dimension Vista Triglyceride assay, patient and QC triglyceride results may be erroneously elevated as a result of contamination.

Distribution Pattern

Nationwide domestic distribution. Worldwide foreign distribution.

Code Information

Millipore lot number: F7MA50339 F7MA56608 F7NA62488 F8AA13966 F8AA93567 F8BA20457 F8BA35774 F8DA59251 F8HA64477 F8HA79819 F8HA91305 F8JA97876 F8MA06374

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1014-2019
openFDA file date
Oct 21, 2026
Loaded by RegDataLab
Oct 23, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.