Z-1018-2022 Class II Terminated

Recalled by Becton Dickinson & Co. — Sparks, MD

Recall Details

Product Type
Devices
Report Date
May 11, 2022
Initiation Date
March 9, 2022
Termination Date
January 19, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
15 systems

Product Description

BD Synapsys Microbiology Informatics Solution, Catalog Number 444150, Software Version 4.10

Reason for Recall

When updating BD Kiestra customers to BD Synapsys version 4.10, the custom antibiotic codes set by the customer are overwritten with the BD default antibiotic codes.

Distribution Pattern

Worldwide distribution US Nationwide distribution in the states of CA, IL, MI, NY, OH, SC, and TN. The countries of Austria, France, Germany, Netherlands, Sweden.

Code Information

All systems running BD Synapsys version 4.10; UDI: 00382904441500; Serial Numbers: SL00530 SL00515 SL00527 SL00502 SL00540 SL00564 SL00549 SL00509 SL00003 SL00004 SL00010 SL00001 SL00005 SL00002 SL000015 SL01010 SL01022