Z-1018-2023 Class II Ongoing

Recalled by Biomet, Inc. — Warsaw, IN

FDA device recall Z-1018-2023 was initiated by Biomet, Inc. on Dec 15, 2022 and is designated Class II. Reason for recall: Packaged in the incorrect shelf carton. Incorrect shelf carton; outer carton is smaller than the required outer carton, which forces the Tyvex pouch to be folded and packaged in the carton causing… The recall status is ongoing. Affected quantity: 7 units.

Recall Details

Product Type
Devices
Report Date
Feb 1, 2023
Initiation Date
Dec 15, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7 units

Product Description

Oxford Partial Knee System, Fixed Lateral Tibial Construct- Intended to replace part of a knee joint when used in conjunction with a femoral and tibial component Item Number: 154341

Reason for Recall

Packaged in the incorrect shelf carton. Incorrect shelf carton; outer carton is smaller than the required outer carton, which forces the Tyvex pouch to be folded and packaged in the carton causing compound creases and compromise the sterile barrier

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of IN, KY and the countries of Canada, New Zealand, Japan, Netherlands.

Code Information

UDI: (01) 05019279515349 (17) 241123 (10) 077830 Lot Number: 077830

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1018-2023
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 24, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.