Z-1020-2022 Class II Ongoing

Recalled by Siemens Healthcare Diagnostics, Inc. — East Walpole, MA

Recall Details

Product Type
Devices
Report Date
May 11, 2022
Initiation Date
March 3, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2,044 units

Product Description

Atellica IM Enhanced Estradiol (eE2) 100T - In vitro diagnostic use in the quantitative determination of estradiol SMN: 10995561 All lots

Reason for Recall

Falsely Elevated Results Observed with Plasma Specimens, could lead a clinician to misinterpret patient results

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Brazil, Bulgaria, Burkina, Faso Country AG, Cyprus, Denmark, Egypt, Estonia, Finland, France, French Polynesia, Georgia, Greece, Guadeloupe, India, Ireland, Israel, Italy, Kazakhstan, Latvia, Lithuania, Macedonia, Malta, Morocco, Nepal, Netherlands, Norway, Palestinian Ter, Paraguay, Poland, Portugal, Qatar, Romania, Russian Fed., Slovakia, Spain, Sweden, Switzerland, Turkey, Turkmenistan, U.A.E., Ukraine, United Kingdom, Vatican city.

Code Information

All lot numbers UDI: (01)00630414598857(10)53376082(17)20220521 (01)00630414598857(10)53377082(17)20220521 (01)00630414598857(10)70533084(17)20220804