Z-1021-2018 Class II Terminated

Recalled by Zimmer Biomet, Inc. — Warsaw, IN

Recall Details

Product Type
Devices
Report Date
March 21, 2018
Initiation Date
November 29, 2017
Termination Date
April 2, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
13,227 in total

Product Description

VERSYS 10 INCH BEADED FC REV 15.0X250MM BWD LT & RT VERSYS BEADED FULLCOAT REVISION HIP PROSTHESIS

Reason for Recall

The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.

Distribution Pattern

Worldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Brazil, Bolivia, Canada, Chile, China, Colombia, Ecuador, El Salvador, India, Japan, Malaysia, Mexico, Netherlands, New Zealand, Nicaragua, Panama, Seoul-Korea, Singapore, Taiwan, Thailand, and UAE.

Code Information

Item Number: 00-7843-015-11 Lot Numbers: 62097294 11006832 62116495 61981821 Item Number: 00-7843-015-12 Lot Numbers: 62085294 11004177 62097244 61759259 62133149 61807097 77000821 61963730