Z-1021-2025 Class II Ongoing

Recalled by B. Braun Medical, Inc. — Allentown, PA

Recall Details

Product Type
Devices
Report Date
February 5, 2025
Initiation Date
December 4, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1090 units

Product Description

Design Options COMBINED SPINAL/EPIDURAL TRAY- Intended for administration of analgesic/anesthetic agent to intrathecal space and epidural space for pain control. Model/Catalog Number: 530159

Reason for Recall

Incorrect catheter connector within the kit resulting in inability to provide a secure connection to the epidural catheter. The incorrect catheter connector is smaller than the required catheter connector and therefore is unable to be used to administer medication.

Distribution Pattern

US Nationwide distribution in the states of MO, OK.

Code Information

UDI-DI 04046964669286 (Primary); 04046964669279 (Unit of Use); Lot Number 0061916572