Z-1021-2025 Class II Ongoing

Recalled by B. Braun Medical, Inc. — Allentown, PA

FDA device recall Z-1021-2025 was initiated by B. Braun Medical, Inc. on Dec 4, 2024 and is designated Class II. Reason for recall: Incorrect catheter connector within the kit resulting in inability to provide a secure connection to the epidural catheter. The incorrect catheter connector is smaller than the required catheter… The recall status is ongoing. Affected quantity: 1090 units.

Recall details

Product Type
Devices
Report Date
Feb 5, 2025
Initiation Date
Dec 4, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1090 units

Product description

Design Options COMBINED SPINAL/EPIDURAL TRAY- Intended for administration of analgesic/anesthetic agent to intrathecal space and epidural space for pain control. Model/Catalog Number: 530159

Reason for recall

Incorrect catheter connector within the kit resulting in inability to provide a secure connection to the epidural catheter. The incorrect catheter connector is smaller than the required catheter connector and therefore is unable to be used to administer medication.

Distribution pattern

US Nationwide distribution in the states of MO, OK.

Code information

UDI-DI 04046964669286 (Primary); 04046964669279 (Unit of Use); Lot Number 0061916572

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1021-2025
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.