Z-2385-2024 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- July 24, 2024
- Initiation Date
- June 17, 2024
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 100 units
Product Description
Infusomat 60 DROP METRISET PUMP SET,3 SFLINE ASV-Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/Catalog Number: 363010
Reason for Recall
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.
Distribution Pattern
Worldwide distribution - US Nationwide and the country of Canada.
Code Information
UDI-DI 04046964294174; All Lots Distributed After 05SEP2023.