Z-1026-2018 Class II Terminated
FDA device recall Z-1026-2018 was initiated by Zimmer Biomet, Inc. on Nov 29, 2017 and is designated Class II. Reason for recall: The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed… The recall status is terminated (terminated Apr 2, 2020). Affected quantity: 13,227 in total.
Recall Details
- Product Type
- Devices
- Report Date
- Mar 21, 2018
- Initiation Date
- Nov 29, 2017
- Termination Date
- Apr 2, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 13,227 in total
Product Description
VERSYS 6 INCH BEADED FC STEM 12X160MM LM 13X160MM LM 15X160MM LM 17X160MM LM Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Reason for Recall
The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.
Distribution Pattern
Worldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Brazil, Bolivia, Canada, Chile, China, Colombia, Ecuador, El Salvador, India, Japan, Malaysia, Mexico, Netherlands, New Zealand, Nicaragua, Panama, Seoul-Korea, Singapore, Taiwan, Thailand, and UAE.
Code Information
Item Number: 00-7843-012-36 Lot Number: 62132189 Item Number: 00-7843-013-36 Lot Numbers: 62085292 62120673 Item Number: 00-7843-015-36 Lot Numbers: 61695988 62089503 Item Number: 00-7843-017-36 Lot Number: 62097290