Z-1027-2018 Class II Terminated

Recalled by Zimmer Biomet, Inc. — Warsaw, IN

Recall Details

Product Type
Devices
Report Date
March 21, 2018
Initiation Date
November 29, 2017
Termination Date
April 2, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
13,227 in total

Product Description

VERSYS 7.5 IN BEADED FC +10 CALCAR 13.5X180MM STR +10 CALCAR 15.5X180MM STR +20 CALCAR 16.5X180MM STR Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented

Reason for Recall

The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.

Distribution Pattern

Worldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Brazil, Bolivia, Canada, Chile, China, Colombia, Ecuador, El Salvador, India, Japan, Malaysia, Mexico, Netherlands, New Zealand, Nicaragua, Panama, Seoul-Korea, Singapore, Taiwan, Thailand, and UAE.

Code Information

Item Number: 00-7843-013-07 Lot Numbers: 61946025 61735805 61993658 61768679 62029060 61815080 62043103 61879441 77000599 Item Number: 00-7843-015-07 Lot Number: 62135320 Item Number: 00-7843-016-17 Lot Numbers: 61879439 61993648