Z-1027-2024 Class II Ongoing
FDA device recall Z-1027-2024 was initiated by Becton Dickinson & Co. on Jan 8, 2024 and is designated Class II. Reason for recall: There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional… The recall status is ongoing. Affected quantity: 2,363,168 total units.
Recall Details
- Product Type
- Devices
- Report Date
- Feb 14, 2024
- Initiation Date
- Jan 8, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2,363,168 total units
Product Description
BD BBL Sensi Disc Ciprofloxacin- 5 ug, used for semi-quantitative in vitro susceptibility testing. Catalog Number 231657¿and 231658¿
Reason for Recall
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
Distribution Pattern
Domestic distribution nationwide. International distribution worldwide.
Code Information
Catalog No. 231657¿ UDI-DI 00382902316572 Lots 1060926 1279842 2028533 2091948 2301402 3214964 3248370; Catalog No. 231658¿ UDI-DI 30382902316580 Lots 1335458 2336475 3062295 3093427 3163048 3214964 3248370 0344330 1007112 1060926 1127037 1279842 1335457 2028533 2028547 2091948 2091980 2301402 2336474