Z-1031-2013 Class II Terminated

Recalled by Medtronic Neuromodulation — Minneapolis, MN

FDA device recall Z-1031-2013 was initiated by Medtronic Neuromodulation on Feb 27, 2013 and is designated Class II. Reason for recall: Medtronic has found through device testing that if the optional footswitch is not connected to the Model 8930 RF generator during use, an electrostatic discharge (i.e. static electricity) to the… The recall status is terminated (terminated Nov 21, 2013). Affected quantity: 848 (696 US, 152 OUS).

Recall Details

Product Type
Devices
Report Date
Apr 10, 2013
Initiation Date
Feb 27, 2013
Termination Date
Nov 21, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
848 (696 US, 152 OUS)

Product Description

Medtronic Prostiva RF Therapy Generator. The Prostiva RF system is indicated for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH) in men over the age of 50 with prostate sizes between 20 and 50 cm3.

Reason for Recall

Medtronic has found through device testing that if the optional footswitch is not connected to the Model 8930 RF generator during use, an electrostatic discharge (i.e. static electricity) to the footswitch connector port may toggle the RF output on or off. If this were to occur, it could result in unintentionally starting or stopping the lesion process, depending on the current status of the equi

Distribution Pattern

Nationwide Distribution including Puerto Rico and all States in continental USA except ID, MT, and NH.

Code Information

all Model 8930 radio frequency (RF) generators are affected

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)