Z-1031-2023 Class II Ongoing
FDA device recall Z-1031-2023 was initiated by MEDLINE INDUSTRIES, LP - Northfield on Nov 17, 2022 and is designated Class II. Reason for recall: Potential for sterile blades, within non-sterile SPT kits containing the blades, to puncture the outer foil layer causing a breach in the sterile packaging. The recall status is ongoing. Affected quantity: 57 cases (114 units).
Recall Details
- Product Type
- Devices
- Report Date
- Feb 8, 2023
- Initiation Date
- Nov 17, 2022
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 57 cases (114 units)
Product Description
Non-sterile procedural trays labeled as LAPAROTOMY CDS, 2 per case
Reason for Recall
Potential for sterile blades, within non-sterile SPT kits containing the blades, to puncture the outer foil layer causing a breach in the sterile packaging.
Distribution Pattern
US Nationwide distribution in the states of AR, MD, WA, WI.
Code Information
Item Number: CDS980894A, UDI/GTIN Case: 40889942401292, UDI/GTIN Each: 10889942401291, Lot Numbers: 22HMD613, 22JMB665